Screenshots
How Much Does Curien Hub Make?
Platform Performance
Rating Distribution
Top Countries
Reviews (1)โผ
Tested multimeter ohm setting and immediately found an issue. Connected leads together, meter reads 1.03 ohms. Hit zero function and meter appears to have zeroed correctly showing 0m ohms. Also shows 1.03 on the zero button. Disconnect the leads for a few seconds and reconnected to test return to zero. The reading returns to 1.03 ohms instead of 0 ohms. To test further I checked a solenoid that two other meters read at just above 20 ohms. This meter read over 21 showing its remaining biased about 1 ohm. I tried a firmware update to possibly correct the issue but the meter will not update, just reads at 0% for over 30 minutes. I wish it worked well but at this point I just donโt trust it.
Version History (2)โผ
Available In (115 Countries)โผ
Supported Languages (1)โผ
App Detailsโผ
This app is to be used with all Curien Bluetooth enabled measurement devices for the central reading, analysis and interpretation of data. This platfo...